Diabetes is a common metabolic endocrine disease, which is caused by the lack of insulin or abnormal receptors in the human body. It is characterized by high blood sugar and is a popular disease worldwide. In recent years, its incidence has shown a significant upward trend, and currently about 10% of adults worldwide suffer from this disease [1]. In China, there are about 40 million diabetic patients. The current treatment method is mainly to regulate the glucose metabolism in the patient. The important basis for clinical treatment administration is the blood glucose content of the patient. Therefore, it is important to track and evaluate the control of diabetes through self-monitoring blood glucose meters. In particular, the self-monitoring blood glucose meter can easily and quickly detect the results in hospitals or even homes. At present, the market blood glucose meter products can only give blood glucose values. It is still difficult for patients to adjust the treatment plan and diet control accordingly. In addition, the test accuracy and test range are insufficient. Therefore, it is of great significance to study high-precision and intelligent blood glucose monitoring instruments.

1 Detection principle

The large-scale detection instruments used in clinical diagnosis are not suitable for small and medium-sized hospitals, emergency departments, and patients for long-term self-monitoring of the disease due to their large size, high price, and cumbersome detection process. The practical application of enzyme electrodes in biosensors has made rapid progress Enzyme electrodes for glucose, lactic acid, cholesterol, urea, amino acids are being widely used in medical testing. A common feature of these sensors is the use of molecular recognition capabilities of biological materials such as microorganisms, enzymes, antibodies, and sensors that use biological materials as molecular recognition elements; and then use the electrons generated by the reaction of the detected substance and enzyme contained in the sample solution to To reduce the electron acceptor, the measuring device uses an electrochemical method, that is, measures the amount of reduction of the electron acceptor, and performs quantitative analysis of the detected substance.

The detection principle of this instrument is based on glucose oxidase (GOD) solidified on the surface of the electrode. When blood drops into the blood glucose test electrode, an oxidation-reduction reaction occurs. During the reaction, the ferrous iron ions in the transmission medium lose electrons and occur. After the oxidation reaction, glucose oxidase oxidizes glucose to produce H 2 O 2 and gluconic acid. H 2 O 2 oxidizes the ferrous iron ions. The electron gain and loss phenomenon occurs during the redox process. Under a certain voltage, an oxidation current is formed. By detecting that the change in current is approximately linear with the glucose concentration, the purpose of detecting the blood glucose concentration is achieved. The specific reaction equation is as follows:

反应方程式

2 Instrument design and implementation

2.1 Hardware structure

The whole instrument system is composed of enzyme electrode sensing part, signal conditioning (current voltage conversion, amplifying and filtering part), temperature compensation part, liquid crystal display part, single chip microcomputer and intelligent chip, as shown in Figure 1. The micro current generated by the blood drop after the enzyme electrode is small and can only reach the level of micro amperes, which is not convenient for measurement and analysis, so it is first converted into a voltage signal, and then the voltage is amplified. Since the system noise generated by the interference signal from the power supply and various factors affects the test accuracy, the filter circuit should be designed to remove the interference signal to make the test more accurate. The processed voltage value is sent to the MSP430 with built-in A / D conversion, and the single-chip computer calculates the blood glucose concentration value, and then displays the result using the liquid crystal.

硬件结构

2.2 Main test process

The signal conditioning circuit in the detection system is shown in Figure 2. After the instrument is powered on, the single-chip microcomputer controls the multi-way switch S1 to close, and the input of the preamplifier is the partial voltage value of R1 and R2, which is set to 0.4V. According to the characteristics of the follower and ideal op amp, the output of the subsequent stage will not change to 0.4V at this time. After A / D conversion, it is input to the single chip microcomputer, and the control display shows "please insert a test strip". When the system inserts a blood glucose test strip, S2 is closed and R4 is introduced into the circuit. The post-stage op amp constitutes an in-phase amplifier, and adjusts R4 so that the circuit output value is 0.6V. After A / D conversion, it is input to the single chip microcomputer, and the display screen is controlled to display "please drop blood". After dripping blood, the current signal is input from "Input", and the output voltage signal is abruptly changed and input into the single chip microcomputer. After the reaction, the blood glucose value is calculated and displayed according to the voltage change.

2.3 Temperature compensation

Changes in ambient temperature cause changes in the zero drift and sensitivity of the detection system and cause measurement errors. In order to eliminate the influence of the ambient temperature, the temperature compensation circuit in the system is realized by the single bus miniature temperature sensor DS18B20 of DALLAS company. Its temperature measurement range is -55 to + 125 ° C, measurement resolution is 0.0625 ° C, and temperature measurement accuracy is ± 0. .5 ° C. The temperature signal is input to the single-chip microcomputer through the multi-way switch, and the CPU automatically corrects the error of the test result according to the temperature characteristic of the blood glucose test electrode.

2.4 Smart chip

In order to further improve the practicality of the instrument and to facilitate the patient as much as possible, the principle of artificial intelligence is introduced into the instrument, and the patient can obtain personal disease information and display simple treatment methods and precautions while detecting their blood glucose concentration. This function is mainly achieved by calling the information in the solidified smart chip with diabetes knowledge base. The knowledge base solidified in the chip is based on the treatment guidelines approved and published by the World Health Organization, the International Diabetes Federation and the Chinese Diabetes Association. The guidelines are authoritative in diabetes analysis. After measuring the blood glucose value, the patient can select the "inference" function on the operation interface.

3. Instrument performance analysis and testing

3.1 Easy operation and low power consumption

The patient only needs to follow the prompts to complete the whole test process after booting, and the operation interface is simple and friendly. According to the output voltage regulated by the test as shown in Figure 2, it can be seen that the response tends to be stable about 20s after the blood is dripped, and the test time of the patient under the correct operation step does not exceed 25s. Because the siphon test paper is used for blood collection, only 3 to 5 μL is required for a single blood collection. Select MSP430 series low-power single-chip microcomputer, the instrument can automatically power off after taking out the test paper according to the voltage change during the test, which can effectively save energy consumption.

3.2 Wide test range

According to the diagnostic criteria for non-insulin-dependent diabetes mellitus (NIDDW) recommended by the WTO in 1998, blood glucose of less than 7 mmol / L is non-diabetic (fasting blood glucose of 3.5 to 7 mm / L is the normal range), and 7 to 11.1 mmol / L. For low glucose tolerance, greater than 11.1mmol / L is diabetic. After testing with different samples, the test range of this instrument can reach 2.2 to 27.8 mm / L, which completely covers the range of possible values ​​of human blood glucose.

3.3 Reliable test data

In order to verify the true reliability of the measurement data of the blood glucose detector, a clinical test was conducted in the research room of the First Hospital of Shanxi University. Respectively test samples with different blood glucose concentrations with an intelligent blood glucose meter and an Italian large-scale biochemical analyzer (BT? 3000). The test results show that the two have a good consistency. The error at high blood glucose is greater than that at low and normal blood glucose, which requires further calibration and calibration. The test values ​​of large-scale biochemical analysis are used as standard values, and the test results of 30 groups are shown in Table 1. Perform regression analysis on the test data, and the fitting results are shown in Figure 3. The regression equation is

公式

In the formula: regression coefficient b = 0.9985, intercept a = -0.0806, correlation coefficient R = 0.996, standard deviation SD = 0.238.

拟合结果

Table 1 Comparison result of smart blood glucose meter and standard blood glucose value

4 Conclusion

The blood glucose concentration measured by the intelligent blood glucose self-monitor has a significant correlation with the measurement result of the high-precision blood glucose meter, and has the advantages of simple operation, short measurement time, more accurate and reliable results, high intelligence, and low power consumption. The improvement can be an ideal instrument for diabetic patients to monitor and control their own blood glucose concentration at home.

940nm LED is one of the most popular IR LED in our life.

It had been widely used in most of night vision camera, remove control ect.


We supply variety of 940nm IR LED(Infrared LED) products. Including Through-hole 940nm LED, SMD 940nm LED and high-power 940nm LED. We can also produce 940nm LED according to your requirement.

There are two kinds of shape for those 940nm led: SMD 940nm LED and Through-hole 940nm LED.


For the Through-hole Light Emitting Diode 940nm LED,

We can customize the shape, the lighting angle, the number of emitting source, the flat pin LED and braided LED. Such as: infrared 5mm 940nm LED with 5 degree, infrared 5mm 940nm LED with 20 degree, infrared 5mm 940nm LED with 30 degree, infrared 5mm 940nm LED with 45 degree, infrared 5mm 940nm LED with 60 degree, infrared 5mm 940nm with 90 degree, infrared 5mm 940nm LED with 120 degree. Infrared 3mm 940nm LED with 3 degree, infrared 3mm 940nm LED with 20 degree, infrared 3mm 940nm LED with 30 degree, infrared 3mm 940nm LED with 45 degree, infrared 3mm 940nm LED with 60 degree, infrared 3mm 940nm with 90 degree, infrared 3mm 940nm LED with 120 degree ect.

There are many other shapes or color of lens for your choose. Customized infrared LED are available


For the SMD LED 940nm LED,

We can supply dual-chip infrared LED, three-chip infrared LED, multi-chip infrared LED, high voltage LED, flashing infrared LED and variety of size SMD LED. For instance: 3528 SMD infrared 940nm LED, 2835 SMD infrared 940nm LED, 3014 SMD infrared 940nm LED, 1206 SMD infrared 940nm LED, 3020 SMD infrared 940nm LED.
There are also have many other shapes to choose, like the 5050 SMD infrared LED, the 5730 SMD infrared LED ect. You can choose any one of them for your requirement.


940nm led is a infrared LED(IR LED). Common infrared LEDs are like: 940nm IR LED, 950nm IR LED and so on.




940nm LED

940nm LED,940nm LED,940nm IR LED,High Brightness 940nm LED

Shenzhen Best LED Opto-electronic Co.,Ltd , https://www.bestsmd.com